. 2002 Nov;40(11):4346–4348. doi: 10.1128/JCM.40.11.4346-4348.2002

FIG. 1.

FIG. 1.

Comparison of HCV load measurements with nested real-time QRT-PCR and Amplicor Monitor assay results. Sixteen HCV-positive serum samples with a wide range of viral loads, measured with the Amplicor Monitor assay and ranging from 1 × 103 to 1.2 × 106 copies/ml, were evaluated with the nested real-time QRT-PCR assay (r = 0.862, P < 0.0001, n = 16).