. 2002 Apr;40(4):1518–1521. doi: 10.1128/JCM.40.4.1518-1521.2002

TABLE 1.

Precision of LCx using the 0.2-ml sample extraction procedure

Specimen (n = 12) Mean copies/ml
SD (% CVa)
Log No. Intrarun Interrun Total
2171 2.62 561 0.36 (13.8) 0.17 (6.5) 0.34 (13.0)
1970 2.95 1,243 0.46 (15.7) 0.07 (2.4) 0.39 (13.1)
Panel 3 3.17 1,772 0.24 (7.5) 0.19 (5.9) 0.30 (9.5)
2199 3.25 2,262 0.25 (7.7) 0.19 (5.7) 0.28 (8.5)
2201 3.94 9,456 0.09 (2.2) 0.15 (3.9) 0.18 (4.5)
2186 4.16 15,577 0.12 (2.9) 0.11 (2.7) 0.17 (4.0)
Panel 4 4.59 39,792 0.06 (1.4) 0.02 (0.5) 0.07 (1.5)
1971 4.77 61,769 0.07 (1.6) 0.09 (2.0) 0.12 (2.6)
2203 5.62 470,242 0.18 (3.1) 0.15 (2.6) 0.23 (4.1)
1974 5.70 536,872 0.17 (2.9) 0.03 (0.5) 0.14 (2.4)
2205 6.62 4,280,114 <0.01 (0.7) 0.08 (1.2) 0.09 (1.4)
a

CV, coefficient of variation.