Table 1.

Mean values ± standard deviation of 20 in vitro Doppler flow measurements of the average peak velocity at each perfusion speed in glass pipes of 1.5 – 5.5 mm. The perfusion speed was adjusted up to 700 ml/min (2.5 – 5.5 mm diameter) and up to 200 ml/min in pipes of 1.5 mm diameter. The flow velocity was calculated of the blood flow (Perf.) and the luminal area of the pipe. The velocity value at the critical Reynolds number (graphic file with name 1476-7120-4-14-i2.gif) is marked for each diameter.

Diameter: 1.5 mm 2.5 mm 3.5 mm 4.5 mm 5.5 mm
Velocity of CRN 125.8 cm/s 75.5 cm/s 53.9 cm/s 41.9 cm/s 34.3 cm/s

Perf. (ml/min) Velocity (cm/s) APV Velocity (cm/s) APV Velocity (cm/s) APV Velocity (cm/s) APV Velocity (cm/s) APV
Mean ± SD Mean ± SD Mean ± SD Mean ± SD Mean ± SD

50 47.2 94.7 ± 1.23 17.0 35.7 ± 0.65 8.7 17.4 ± 0.48 5.2 11.4 ± 0.48 3.5 7.1 ± 0.08
100 94.3 183.7 ± 1.56 34.0 68.0 ± 0.63 17.3 34.1 ± 0.97 10.5 21.6 ± 0.73 7.0 14.0 ± 0.0

150 141.5 259.6 ± 3.88 50.9 97.0 ± 1.52 26.0 51.6 ± 0.73 15.7 32.6 ± 0.66 10.5 20.7 ± 0.57
200 188.6 328.9 ± 7.56 67.9 120.5 ± 2.38 34.7 67.1 ± 0.67 21.0 42.3 ± 0.71 14.0 27.2 ± 0.51

250 84.9 145.2 ± 3.79 43.3 82.1 ± 0.77 26.2 51.7 ± 0.78 17.5 33.5 ± 0.87
300 101.9 169.6 ± 5.28 52.0 96.5 ± 0.97 31.4 62.3 ± 0.78 21.1 39.1 ± 1.56

350 118.8 195.3 ± 6.91 60.6 110.5 ± 0.80 36.7 70.9 ± 0.73 24.6 44.5 ± 2.36

400 135.8 220.8 ± 8.29 69.3 124.3 ± 1.61 41.9 82.3 ± 0.71 28.1 50.8 ± 2.58
450 152.8 246.7 ± 11.04 78.0 140.7 ± 0.90 47.2 87.1 ± 1.34 31.6 55.6 ± 3.41

500 169.8 271.1 ± 12.50 86.6 152.1 ± 1.51 52.4 96.4 ± 1.96 35.1 61.9 ± 2.74
550 186.7 296.5 ± 14.73 95.3 163.6 ± 1.62 57.6 105.5 ± 1.57 38.6 67.1 ± 3.63
600 203.7 340.6 ± 14.79 103.9 178.8 ± 2.32 62.9 115.6 ± 1.83 42.1 73.4 ± 3.98
650 220.7 362.6 ± 15.40 112.6 192.7 ± 1.56 68.1 124.8 ± 1.96 45.6 79.5 ± 4.10
700 237.7 368.3 ± 25.65 121.3 206.9 ± 2.08 73.4 133.8 ± 2.32 49.1 85.8 ± 4.91