TABLE 1.

Comparative distribution of EIA results for United Kingdom blood donor and Ghanaian child samples tested with VP2 genotype 1 and 3 antigens

VP2G1Cam test result No. of samples with VP2G3Cam test result
No. of samples discordant/total no. of samples
Positive Negative EQa
United Kingdom donorsb
    Positive 271 0 2 2/273
    Negative 0 125 0 0/125
    EQ 0 1 1 1/2
    No. of samples discordant/total no. of samples 0/271 1/126 2/3 3/400
Ghanaian childrenc
    Positive 30 1 2 3/33
    Negative 0 359 2 2/361
    EQ 2 1 3 3/6
    No. of samples discordant/total no. of samples 2/32 2/361 4/7 8/400
a

EQ, equivocal.

b

Agreement of 99.3% (95% CI, 97.8 to 99.8) between samples tested with VP2G3Cam and VP2G1Cam was observed. Kappa and Spearman correlation coefficients were 0.98 and 0.97, respectively.

c

Agreement of 98.0% (95% CI, 96.1 to 99.1) between samples tested with VP2G3Cam and VP2G1Cam was observed. Kappa and Spearman correlation coefficients were 0.89 and 0.77, respectively.