. 2003 Jan;10(1):78–82. doi: 10.1128/CDLI.10.1.78-82.2003

TABLE 3.

Discrepant and divergent results obtained with the BBI reference panela

Sample no. VCA IgG/VCA IgM/EBNA IgG score
BBI reference Novitec Biotest Virotech DiaSorin IFA
1 −/−/− +/−/+
2 −/−/− +/−/− −/−/+
3, 4, 5 ±/−/−b −/−/− −/−/− −/−/− −/−/− −/−/−
6 +/+/−c +//−
7 +/+/−c +/+/+
8 +/+/−c +//− +/+/+
9 +/+/±c,d +/+/+ +/+/+ +/+/+ +/+/− +//+
10, 11 +/+/−c
12 +/±/+e +/+/+ +/+/+ +/−/ +/−/+ +/−/+
13 +/±/+e +/−/+ +/−/+ +/+/+ +/−/+ +/−/+
14 +/±/−e +/−/+ +/+/+ +/−/− +/−/+ +/−/−
15 +/−/±d +/−/− +/−/+ +/−/− +/−/+ +/−/−
16 +/−/±d +/−/+ +/−/+ +/−/+ +/−/− +/−/−
17 +/−/−f +/−/− +/−/+ /−/+ +/−/+ +/−/+
18, 19, 22 +/−/+g
20, 21 +/−/+ +/+/+
23 +/−/±h +/−/+ +/−/+ +/−/− +/−/+ +/−/+
24 +/−/±d /−/+ +/−/− +/−/+ /−/−
25 +/−/±d +/−/+ +/+/+ +/−/− +/−/+ +/−/−
a

The BBI panel was pretested with three assays for heterophilic antibodies, two EIAs, and two IFAs (see Materials and Methods). If not otherwise indicated, samples were negative for heterophilic antibodies. Only discrepant results and all samples with divergent results within the reference assays (as indicated by indices) are shown. Discordant results are in bold. +, positive; −, negative; ±, divergent.

b

The Gull IFA produced a weak positive result with sample 5.

c

All three assays for heterophilic antibodies yielded positive results.

d

Zeus and Gull IFA negative.

e

Meridian EIA positive.

f

False negative by Meridian VCA-IgG EIA since the respective Meridian assay was EBNA-1 IgG positive.

g

Heterophilic antibodies indicated by weakly positive Seradyn assay result obtained with sample 22.

h

Incstar EIA negative.