Published in final edited form as: Rapid Commun Mass Spectrom. 2003;17(7):637–645. doi: 10.1002/rcm.960

Table 2.

Repeatability and intra-day and inter-day precision of the LC/APCI-MS/MS method, calculated at 25 nM. Results are expressed as % relative standard deviation (% RSD)

Compound Repeatabilitya % Intra-dayb % Inter-day %c
MDA 1.3 5.2 6.9
Acrolein 2.7 3.5 7.5
4-HHE 2.6 4.4 6.9
4-HNE 2.1 2.6 7.8
n-hexanal 2.2 4.7 7.4
n-heptanal 4.4 7.9 9.9
n-nonanal 3.6 6.2 8.5
a

n = 12.

b

n = 24; four derivatized samples, each injected twice, at the beginning, in the middle and at the end of the working session.

c

n = 24; solutions of aldehydes were prepared daily from the 5 mM solution in acetonitrile and derivatized.