Skip to main content
. Author manuscript; available in PMC: 2006 May 18.
Published in final edited form as: Biomaterials. 2005 Nov;26(33):6657–6663. doi: 10.1016/j.biomaterials.2005.04.026

Fig. 4.

Fig. 4

The wear rates of accelerated aged 0.1 and 0.3 wt% α-tocopherol-blended UHMWPE. Accelerated aging was done at 80 °C for 5 weeks in air. For comparison, 127-kGy irradiated and α-tocopherol-diffused UHMWPE [6], 38-kGy sterilized and 100-kGy irradiated and melted GUR 1050 [14] are also included. Accelerated aging for the 38 and 100-kGy irradiated samples taken from Muratoglu et al. [14] were done at 80 °C for 3 weeks in air.