Table 3.

Reliability data of the method for the determination of eicosanoids in exhaled breath condensate.

Analyte Intra-day precision Inter-day precision Accuracy Calibration (Y = ax+b)
Q1 Q2 Q1 Q2 relative recovery (%) a (x10e-3) b (x10e-3)
RSD(%) RSD(%) RSD(%) RSD(%) mean range
8-iso-PGF 2alfa 5 5 16 11 94 88–107 0,098 2,21
11-β-PGF 2alfa 4 2 10 8 102 97–108 0,084 0,358
PGF 2alfa 2 2 13 8 102 99–104 0,254 0,314
PGE 2 4 2 5 6 99 96–107 0,628 1,92
PGD 2 4 2 4 6 102 94–112 0,345 0,089
13,14-dihydro-15-keto-PGE 2 2 2 8 8 94 90–100 1,46 -0,245
13,14-dihydro-15-keto-PGD 2 2 3 9 8 104 99–110 1,04 4,88
LTE 4 6 4 10 7 115 96–133 0,238 1,3
Delta 12-PGJ 2 5 5 13 11 105 98–120 1,09 1,2
PGJ 2 5 6 4 8 110 102–119 0,213 1,17
LTB 4 5 5 9 11 110 97–117 0,19 1,4
15-desoxy-delta12,14-PGJ 2 5 5 16 12 159 135–179 0,739 5,46

Q1 and Q2 represents the low and the high concentration level respectively, with 50 pg/mL and 500 pg/mL each analyte.