Skip to main content
. 2006 Jun 21;34(10):e75. doi: 10.1093/nar/gkl364

Table 2.

Rct valuesa, Rct change values and change ratio of Rct values before and after interaction MutS with heteroduplex and homoduplex

GT CT CC AT
Rct (Ω)
    MutS (2.43 ± 0.51) × 104 (3.06 ± 0.43) × 104 (3.82 ± 0.68) × 104 (3.78 ± 0.70) × 104
    2 µM (3.24 ± 0.23) × 104 (3.78 ± 0.18) × 104 (4.12 ± 0.27) × 104 (3.87 ± 0.15) × 104
    6 µM (4.52 ± 0.31) × 104 (4.34 ± 0.25) × 104 (4.42 ± 0.29) × 104 (3.22 ± 0.17) × 104
    10 µM (5.86 ± 0.36) × 104 (4.85 ± 0.24) × 104 (4.92 ± 0.25) × 104 (3.98 ± 0.19) × 104
ΔRct (Ω)
    2 µM 8050 7190 3000 860
    6 µM 20 810 12 730 6000 1410
    10 µM 34 200 17 870 11 020 2030
%(ΔRct)
    2 µM 33 23 8 2
    6 µM 85 41 16 4
    10 µM 140 58 21 6
Detection limitb (µM)  0.044 0.13 0.31 1.56

aThe values of Rct are mean ± SD of three repeated experiments.

bDetection limit Cm=(SmS¯bl)/m, Sm=S¯bl+ksbl(S¯bl, mean of blank; Sbl, standard deviation of blank; k, detection limit constant, blank contains only the MutS solution).