NucliSens HIV-1 QT assay precision estimates

HIV-1 RNA stock dilutiona Nominal mean value (log10 copies/ml) No. of specimens that tested positive Observedb mean value (log10 copies/ml) Overall SDc (log10 copies/ml)
2 6.43 72 6.52 0.14
5 5.53 72 5.54 0.09
8 4.63 72 4.66 0.12
12 3.42 72 3.52 0.18
14 2.82 72 2.93 0.29
15 2.52 72 2.66 0.28
16 2.21 69 2.26 0.36
17 1.91 51 1.98 0.34
a

Serial dilutions were made from a well-characterized HIV-1 RNA stock (titer = 5.4 × 108 HIV-1 RNA copies/ml) as described in Materials and Methods.

b

Observed mean value obtained for each dilution derived from replicate tests per sample.

c

Overall value for all 552 replicate tests, 0.225.