Skip to main content
. 2003 Jan;41(1):242–249. doi: 10.1128/JCM.41.1.242-249.2003

TABLE 2.

Sensitivities, specificities, and predictive values for Directigen rapid antigen immunoassay and LightCycler PCR assay compared to the results of culture for detection of GAS from throat swabsa

Type of assay Sensitivity (%) Specificity (%) Positive predictive value (%) Negative predictive value (%)
Directgen antigen immunoassay 55 (41, 68)b 99 (98, 100) 97 (83, 100) 93 (90, 95)
LightCycler PCR assay 93 (82, 98)b 98 (96, 99) 88 (77, 95) 99 (97, 100)
a

The prevalence rate of positive cultures was 14.3%. Values in parentheses are 95% confidence intervals.

b

Signficantly different by McNemar's test (P < 0.001).