Skip to main content
. 2003 Mar;10(2):303–307. doi: 10.1128/CDLI.10.2.303-307.2003

TABLE 2.

Sensitivity and specificity for Rapid Check HIV 1&2 and Determine tests for both plasma and whole-blood samples

Assay Result Plasma samples
Whole-blood samples
No. HIV positive (n = 137) No. HIV negative (n = 99) % Sensitivity (% Specificity) No. HIV positive (n = 111) No. HIV negative (n = 50) % Sensitivity (% Specificity)
Rapid Check HIV 1&2 Positive 137 1 100 (98.99) 111 0 100 (100)
Negative 0 98 0 50
Determine Positive 135 3 98.54 (96.97) 111 0 100 (100)
Negative 2 96 0 50