. 1979 Mar;54(3):184–189. doi: 10.1136/adc.54.3.184

Clinical evaluation and comparison of the Infrasonde, Arteriosonde, and mercury sphygmomanometer in measurement of blood pressure in children.

J M Savage, M J Dillon, J F Taylor
PMCID: PMC1545239  PMID: 434903

Abstract

Both systolic and diastolic pressures can be measured in children with the mercury sphygmomanometer, the Arteriosonde and the Infrasonde. Estimates made blindly with these instruments were compared with directly measured intra-arterial blood pressures in 50 children aged between 4 days and 14 years. Systolic and diastolic estimates with the three techniques showed highly significant correlations with simultaneous intra-arterial measurements (P less than 0.001). The Infrasonde diastolic estimates were least satisfactory and the slope of the regression line against the intra-arterial pressure differed significantly from unity (y=0.54x+29.53). In 11 small children a satisfactory diastolic estimate could not be obtained with the mercury sphygmomanometer. While the mercury sphygmomanometer should remain the standard hospital equipment, an Arteriosonde would be valuable if it is difficult to hear Korotkoff's sounds on auscultation and if a diastolic pressure is required. For research investigations into childhood blood pressure an Arteriosonde or mercury sphygmomanometer, coupled with a device to exclude observer bias, is probably most suitable. Although the Infrasonde is not sufficiently accurate for research purposes, it is acceptable for routine ward use.

Full text

PDF
185

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Edwards R. C., Goldberg A. D., Bannister R., Raftery E. B. The infrasound blood-pressure recorder. A clinical evaluation. Lancet. 1976 Aug 21;2(7982):398–400. doi: 10.1016/s0140-6736(76)92410-7. [DOI] [PubMed] [Google Scholar]
  2. Elseed A. M., Shinebourne E. A., Joseph M. C. Assessment of techniques for measurement of blood pressure in infants and children. Arch Dis Child. 1973 Dec;48(12):932–936. doi: 10.1136/adc.48.12.932. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. HOLLAND W. W., HUMERFELT S. MEASUREMENT OF BLOOD-PRESSURE: COMPARISON OF INTRA-ARTERIAL AND CUFF VALUES. Br Med J. 1964 Nov 14;2(5419):1241–1243. doi: 10.1136/bmj.2.5419.1241. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Hernandez A., Goldring D., Hartmann A. F., Jr Measurement of blood pressure in infants and children by the Doppler ultrasonic technique. Pediatrics. 1971 Nov;48(5):788–794. [PubMed] [Google Scholar]
  5. Lauer R. M., Rames L. K., Clarke W. R. Blood pressure and its significance in childhood. Postgrad Med J. 1978 Mar;54(629):206–211. doi: 10.1136/pgmj.54.629.206. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Loggie J. M. Systemic hypertension in children and adolescents: causes and treatment. Pediatr Clin North Am. 1971 Nov;18(4):1273–1310. doi: 10.1016/s0031-3955(16)32640-2. [DOI] [PubMed] [Google Scholar]
  7. Londe S., Bourgoignie J. J., Robson A. M., Goldring D. Hypertension in apparently normal children. J Pediatr. 1971 Apr;78(4):569–577. doi: 10.1016/s0022-3476(71)80457-2. [DOI] [PubMed] [Google Scholar]
  8. Moss A. J. Indirect methods of blood pressure measurement. Pediatr Clin North Am. 1978 Feb;25(1):3–14. doi: 10.1016/s0031-3955(16)33527-1. [DOI] [PubMed] [Google Scholar]
  9. Whyte R. K., Elseed A. M., Fraser C. B., Shinebourne E. A., de Swiet M. Assessment of Doppler ultrasound to measure systolic and diastolic blood pressures in infants and young children. Arch Dis Child. 1975 Jul;50(7):542–544. doi: 10.1136/adc.50.7.542. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Archives of Disease in Childhood are provided here courtesy of BMJ Publishing Group

RESOURCES