Skip to main content
. 2002 Dec;40(12):4652–4658. doi: 10.1128/JCM.40.12.4652-4658.2002

TABLE 1.

Accuracy error and CV of standard plasmid pVU58 quantification

Copy no. Accuracy error (%), mean ± SE CV (%)
Intra-assay
Inter-assay
Ct Copies Ct Copies
101 −14.3 ± 4.7 1.21 29.4 1.05 44.4
102 −13.1 ± 2.89 0.5 14.7 1.08 20.6
103 −16.2 ± 2.39 0.72 7.1 0.9 17.6
104 −12.2 ± 2.71 0.68 6.1 1.09 22.8
105 1.8 ± 2.82 0.57 7.3 1.17 20
106 5.7 ± 1.46 0.31 4.8 0.83 9.2