Skip to main content
. 2002 Dec;40(12):4652–4658. doi: 10.1128/JCM.40.12.4652-4658.2002

TABLE 2.

Quantification of HHV-8 DNA with and without calibrationa

Genome equivalents/sample Genome equivalents, mean of recovery ± SD
Intersample CVa
Accuracy error (%)
Uncalibrated Calibrated Uncalibrated Calibrated Uncalibrated Calibrated
800 404 ± 80 677 ± 129 19.8 19 49.5 15.4
300,000 149,860 ± 60,670 282,040 ± 15,500 41.3 8.7 50 6
a

Mean of recovery, inter-sample CV, and accuracy error for different amounts of viral DNA spiked into HHV-8-seronegative human plasma.