Published in final edited form as: IEEE Trans Ultrason Ferroelectr Freq Control. 2006 Jun;53(6):1130–1137. doi: 10.1109/tuffc.2006.1642511

Table I.

Comparison of correlation, error, relative expansion and decrease in diameter for two transmitted pressures and microbubbles with an initial diameter between 1.7 and 3.3 microns. * The signal to noise ratio did not permit accurate estimation of the echo correlation at the lower pressure.

Transmission pressure (kPa) R-T correlation between theory and experiment based on (1) R-T error (μm) between theory and experiment based on (2) Echo correlation between theory and experiment based on (1) Relative expansion (Dmax/D0) Decrease in diameter (%) during pulse
330 (n=15) 0.76 ± 0.05 0.13 ± 0.02 n/a* 1.5 ± 0.3 14 ± 6
620 (n=15) 0.70 ± 0.08 0.29 ± 0.04 - 2.0 ± 0.2 36 ± 9
P value 0.03 2e-13 - 4e-6 1e-8