Table 1.
Comparisons of pulse oximetry measurements on the tongue with the original and refitted oximetry probes
Concentration of inspiratory oxygen (%) | n | Oxygen saturation (%) | Correlation coefficient (r) | |
SpO2refit | SpO2origin | |||
100 | 10 | 100 | 100 | 1.0 |
21 | 10 | 93.2 ± 2.4 (92–96) | 93.4 ± 2.7 (91–96) | 0.95 |
10 | 10 | 81.5 ± 2.2 (77–84) | 81.1 ± 2.5 (78–85) | 0.94 |
Values are means ± SEM (range). SpO2origin, pulse oximetry obtained with the original oximetry probe; SpO2refit, pulse oximetry obtained with the refitted oximetry probe.