. 2003 Jun;41(6):2358–2366. doi: 10.1128/JCM.41.6.2358-2366.2003

TABLE 3.

Performance characteristics of ParaSight F+V prototypes for detection of P. falciparuma

Performance parameter FV98
FV99-1
FV99-2
No. of specimens % of specimens (CI95)
No. of specimens % of specimens (CI95)
No. of specimens % of specimens (CI95)
Peru Thailand Overall Peru Thailand Overall Peru Thailand Overall
Sensitivity
    Any level of parasitemia >0/μl 542 95 (89-98) 99 (98-100) 98 (97-100) 49 92 (75-100) 94 (89-98) 94 (89-97) 164 95 (82-100) 98 (94-100) 98 (94-100)
    >0-500/μl 79 100 (74-100) 97 (90-100) 97 (91-100) 33 83 (36-100) 74 (54-89) 76 (58-89) 24 67 (22-96) 89 (65-99) 83 (63-95)
    501-1,000/μl 31 80 (44-97) 100 (84-100) 94 (79-100) 10 100 (3-100) 100 (66-100) 100 (69-100) 8 100 (3-100) 100 (59-100) 100 (63-100)
    1,001-5,000/μl 102 100 (86-100) 99 (93-100) 99 (95-100) 22 100 (40-100) 100 (81-100) 100 (85-100) 32 100 (74-100) 100 (83-100) 100 (89-100)
    >5,000/μl 330 94 (86-98) 100 (99-100) 99 (97-100) 84 93 (68-100) 100 (95-100) 99 (94-99) 100 100 (81-100) 100 (96-100) 100 (96-100)
Specificity 2,349 35 (31-38) 17 (15-19) 22 (20-24) 86 83 (79-87) 83 (80-86) 83 (80-86) 694 97 (95-99) 89 (85-92) 93 (91-95)
a

Mixed infections (P. falciparum and P. vivax) detected by microscopy were excluded from calculations of device performance (1998, n = 31; 1999, n = 19).