Skip to main content
. 2006 Aug;44(8):2714–2720. doi: 10.1128/JCM.00443-06

TABLE 2.

Sensitivity of NV GGI RT-PCR with and without MS2 IC

;2;;>Spiking titer (no. of NV GGI copies/ml) ;2;;>Corresponding no. of NV GGI copies/PCR RT-PCR without MS2 IC
RT-PCR with MS2 IC
No. of subjects No. of observed responses Probita No. of subjects No. of observed responses Probita
2.47 × 105 4,960 12 12 1.000 12 12 1.000
1.24 × 105 2,480 12 12 1.000 12 12 1.000
6.18 × 104 1,240 12 12 0.999 12 12 0.999
3.09 × 104 620 12 10 0.872 12 10 0.902
1.54 × 104 310 12 7 0.538 12 8 0.541
7.72 × 103 77 12 5 0.246 12 5 0.214
3.86 × 103 39 12 3 0.207 12 2 0.174
1.93 × 103 19 12 2 0.189 12 1 0.155
9.65 × 102 10 12 0 0.181 12 0 0.147
a

Predicted proportion of replicates positive from probit regression analysis (SPS5).