Published in final edited form as: J Biomech Eng. 2005 Aug;127(4):692–699. doi: 10.1115/1.1933949

Table 3.

Femoral and tibial absolute precision results reported by knee fluoroscopy studies in the literature

Reference Fluoroscopy method Knee type Models matched In-plane Trans (mm) Out-of-plane trans (mm) All rots (deg)
Present study Single-plane Natural Bones 0.42 5.6 1.3
Kanisawa et al. [19] Single-plane Natural Bones 1.2 4.0 0.8
Komistek et al. [20] Single-plane Natural Bones 0.45 · · · 0.66
Banks and Hodge [11] Single-plane Artificial Implants 0.21 3.9 1.3
Hoff et al. [13]a Single-plane Artificial Implants 0.46 2.2 0.35
Mahfouz et al. [15] Single-plane Artificial Implants 0.09 1.4 0.40
Tashman et al. [18] Bi-plane Natural Markers 0.06 0.06 0.31
You et al. [16]a Bi-plane Natural Bones 0.23 0.23 1.2
Kaptein et al. [23] Bi-plane Artificial Markers 0.05 0.08 0.07
Kaptein et al. [23] Bi-plane Artificial Implants 0.06 0.14 0.17
a

RMS errors rather than precision results reported