Skip to main content
Archives of Disease in Childhood logoLink to Archives of Disease in Childhood
. 1974 Mar;49(3):213–218. doi: 10.1136/adc.49.3.213

Reliability of transcutaneous monitoring of arterial Po2 in newborn infants

Renate Huch, Dietrich W Lübbers, Albert Huch
PMCID: PMC1648677  PMID: 4825627

Abstract

Blood Po2 was measured transcutaneously on skin `arterialized' by means of a heated Po2 electrode. Oxygen diffuses through the relatively thin skin of newborn infants from the outer capillary layer to an electrode fixed to the skin. Arterial changes were registered with approximately a 10-second delay. 41 comparative transcutaneous and arterial Po2 measurements were performed. Transcutaneous Po2 values correlated quite well with arterial values within the whole Po2 range, the correlation being particularly satisfactory in the Po2 range below 100 mmHg. The method is suitable for long-term monitoring of newborn infants, as it is without risk.

Full text

PDF
218

Images in this article

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. BRADLEY A. F., SEVERINGHAUS J. W., STUPFEL M. Effect of temperature on PCO2 and PO2 of blood in vitro. J Appl Physiol. 1956 Sep;9(2):201–204. doi: 10.1152/jappl.1956.9.2.201. [DOI] [PubMed] [Google Scholar]
  2. Beutnagel H., Gauch D., Fabel H. Der arterielle Sauerstoffpartialdruck beim Neugeborenen in den ersten Lebensminuten. Ein Beitrag zur kontinuierlichen pO 2 -Messung. Z Geburtshilfe Perinatol. 1972 Apr;176(2):117–124. [PubMed] [Google Scholar]
  3. Cochran W. D., Davis H. T., Smith C. A. Advantages and complications of umbilical artery catheterization in the newborn. Pediatrics. 1968 Nov;42(5):769–777. [PubMed] [Google Scholar]
  4. Evans N. T., Naylor P. F. The systemic oxygen supply to the surface of human skin. Respir Physiol. 1967 Aug;3(1):21–37. doi: 10.1016/0034-5687(67)90020-5. [DOI] [PubMed] [Google Scholar]
  5. Harris T. R., Nugent M. Continuous arterial oxygen tension monitoring in the newborn infant. J Pediatr. 1973 Jun;82(6):929–939. doi: 10.1016/s0022-3476(73)80419-6. [DOI] [PubMed] [Google Scholar]
  6. Huch A., Huch R., Arner B., Rooth G. Continuous transcutaneous oxygen tension measured with a heated electrode. Scand J Clin Lab Invest. 1973 May;31(3):269–275. doi: 10.3109/00365517309082431. [DOI] [PubMed] [Google Scholar]
  7. Huch A., Huch R. New method for arterial blood sampling in infants and adults. Arch Dis Child. 1973 Dec;48(12):982–983. doi: 10.1136/adc.48.12.982. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Huch A., Huch R., Rooth G. Monitoring the intravascular Po2 in newborn infants. J Perinat Med. 1973;1(1):53–59. doi: 10.1515/jpme.1973.1.1.53. [DOI] [PubMed] [Google Scholar]
  9. Huch R., Huch A., Lübbers D. W. Transcutaneous measurement of blood Po2 (tcPo2) -- Method and application in perinatal medicine. J Perinat Med. 1973;1(3):183–191. doi: 10.1515/jpme.1973.1.3.183. [DOI] [PubMed] [Google Scholar]
  10. Huch R., Lübbers D. W., Huch A. Quantitative continuous measurement of partial oxygen pressure on the skin of adults and new-born babies. Pflugers Arch. 1972;337(3):185–198. doi: 10.1007/BF00586844. [DOI] [PubMed] [Google Scholar]
  11. Parker D., Key A., Davies R., Scopes J. W., Marcovitch H. A disposable catheter-tip transducer for continuous measurement of blood oxygen tension in vivo. Biomed Eng. 1971 Jul;6(7):313–317. [PubMed] [Google Scholar]
  12. Veasy L. G., Clark J. S., Jung A. L., Jenkins J. L. A system for computerized automated blood gas analysis. Its use in newborn infants with respiratory distress. Pediatrics. 1971 Jul;48(1):5–17. [PubMed] [Google Scholar]

Articles from Archives of Disease in Childhood are provided here courtesy of BMJ Publishing Group

RESOURCES