Skip to main content
. 2003 Jul;41(7):3361–3367. doi: 10.1128/JCM.41.7.3361-3367.2003

TABLE 1.

Numbers and percentages of patient samples in which HIV-1 RNA was detected with AMPLICOR HIV-1 Monitor Test version 1.5, Nuclisens HIV-1 QT, and Quantiplex HIV RNA 3.0 (bDNA) tests

Genetic subtype (gag/env)a No. of samples tested No. (%) of samples in which HIV-1 RNA was detected by:
AMPLICOR HIV-1 Monitor Test version 1.5 Quantiplex 3.0 HIV RNA (bDNA) Nuclisens HIV-1 QT
A1/A1 4 4 (100) 4 (100) 4 (100)
B/B 6 6 (100) 6 (100) 4 (67)
−/B 9 9 (100) 9 (100) 5 (56)
B/− 5 5 (100) 5 (100) 5 (100)
F1/F1 4 4 (100) 4 (100) 4 (100)
G/G 17 17 (100) 17 (100) 3 (18)
B/G 3 3 (100) 3 (100) 0 (0)
G/B 6 6 (100) 6 (100) 1 (17)
U 10 10 (100) 10 (100) 7 (70)
Total 64 64 (100) 64 (100) 33 (52)
a

−/B, genotyped only in the env gene; B/−, genotyped only in the gag gene; U, unknown genotype.