Skip to main content
. 2003 Jul;41(7):3361–3367. doi: 10.1128/JCM.41.7.3361-3367.2003

TABLE 2.

Numbers and percentages of discordant results between AMPLICOR HIV-1 Monitor Test version 1.5, Nuclisens HIV-1 QT, and Quantiplex HIV RNA 3.0 (bDNA) tests

Genetic subtype No. (%) of discordant resultsa/total no. of results obtained with viral load tests:
Quantiplex vs. Amplicor Amplicor vs. Nuclisens Nuclisens vs. Quantiplex
A1 2/4 (50) 4/4 (100) 0/4 (0)
B 6/20 (30) 6/14 (43) 5/14 (36)
F1 0/4 (0) 0/4 (0) 0/4 (0)
G 13/17 (76) 4/5 (80) 4/5 (80)
U 4/10 (40) 2/7 (29) 2/7 (29)
B/G 4/9 (44) 2/4 (50) 2/4 (50)
Total 29/64 (45) 18/38 (47) 13/38 (34)
a

Results for which the difference in log10 HIV-1 RNA copies per ml was >0.5.