. 2001 Sep;85(3):263–267. doi: 10.1136/adc.85.3.263

Measurement of body fat using leg to leg bioimpedance

R Sung 1, P Lau 1, C Yu 1, P Lam 1, E Nelson 1
PMCID: PMC1718893  PMID: 11517118

Abstract

AIMS—(1) To validate a leg to leg bioimpedance analysis (BIA) device in the measurement of body composition in children by assessment of its agreement with dual energy x ray absorptiometry (DXA) and its repeatability. (2) To establish a reference range of percentage body fat in Hong Kong Chinese children.
METHODS—Sequential BIA and DXA methods were used to determine body composition in 49 children aged 7-18 years; agreement between the two methods was calculated. Repeatability for the BIA method was established from duplicate measurements. Body composition was then determined by BIA in 1139 girls and 1243 boys aged 7-16 years, who were randomly sampled in eight local primary and secondary schools to establish reference ranges.
RESULTS—The 95% limits of agreement between BIA and DXA methods were considered acceptable (−3.3 kg to −0.5 kg fat mass and −3.9 to 0.6% body fat). The percentage body fat increased with increasing age. Compared to the 1993 Hong Kong growth survey, these children had higher body mass index. Mean (SD) percentage body fat at 7years of age was 17.2% (4.4%) and 14.0% (3.4%) respectively for boys and girls, which increased to 19.3% (4.8%) and 27.8% (6.3%) at age 16.
CONCLUSION—Leg to leg BIA is a valid alternative method to DXA for the measurement of body fat. Provisional reference ranges for percentage body fat for Hong Kong Chinese children aged 7-16 years are provided.



Full Text

The Full Text of this article is available as a PDF (124.4 KB).

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Bland J. M., Altman D. G. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999 Jun;8(2):135–160. doi: 10.1177/096228029900800204. [DOI] [PubMed] [Google Scholar]
  2. Chan Y. L., Leung S. S., Lam W. W., Peng X. H., Metreweli C. Body fat estimation in children by magnetic resonance imaging, bioelectrical impedance, skinfold and body mass index: a pilot study. J Paediatr Child Health. 1998 Feb;34(1):22–28. doi: 10.1046/j.1440-1754.1998.00147.x. [DOI] [PubMed] [Google Scholar]
  3. Cole T. J., Green P. J. Smoothing reference centile curves: the LMS method and penalized likelihood. Stat Med. 1992 Jul;11(10):1305–1319. doi: 10.1002/sim.4780111005. [DOI] [PubMed] [Google Scholar]
  4. Deurenberg P., Weststrate J. A., Paymans I., van der Kooy K. Factors affecting bioelectrical impedance measurements in humans. Eur J Clin Nutr. 1988 Dec;42(12):1017–1022. [PubMed] [Google Scholar]
  5. Deurenberg P., Yap M., van Staveren W. A. Body mass index and percent body fat: a meta analysis among different ethnic groups. Int J Obes Relat Metab Disord. 1998 Dec;22(12):1164–1171. doi: 10.1038/sj.ijo.0800741. [DOI] [PubMed] [Google Scholar]
  6. Ellis K. J., Shypailo R. J., Abrams S. A., Wong W. W. The reference child and adolescent models of body composition. A contemporary comparison. Ann N Y Acad Sci. 2000 May;904:374–382. doi: 10.1111/j.1749-6632.2000.tb06486.x. [DOI] [PubMed] [Google Scholar]
  7. Ellis K. J., Shypailo R. J., Pratt J. A., Pond W. G. Accuracy of dual-energy x-ray absorptiometry for body-composition measurements in children. Am J Clin Nutr. 1994 Nov;60(5):660–665. doi: 10.1093/ajcn/60.5.660. [DOI] [PubMed] [Google Scholar]
  8. Eston R. G., Cruz A., Fu F., Fung L. M. Fat-free mass estimation by bioelectrical impedance and anthropometric techniques in Chinese children. J Sports Sci. 1993 Jun;11(3):241–247. doi: 10.1080/02640419308729991. [DOI] [PubMed] [Google Scholar]
  9. Gallagher M., Walker K. Z., O'Dea K. The influence of a breakfast meal on the assessment of body composition using bioelectrical impedance. Eur J Clin Nutr. 1998 Feb;52(2):94–97. doi: 10.1038/sj.ejcn.1600520. [DOI] [PubMed] [Google Scholar]
  10. Gutin B., Litaker M., Islam S., Manos T., Smith C., Treiber F. Body-composition measurement in 9-11-y-old children by dual-energy X-ray absorptiometry, skinfold-thickness measurements, and bioimpedance analysis. Am J Clin Nutr. 1996 Mar;63(3):287–292. doi: 10.1093/ajcn/63.3.287. [DOI] [PubMed] [Google Scholar]
  11. Haines A., Heath I., Smith R. Joining together to combat poverty. Everybody welcome and needed. BMJ. 2000 Jan 1;320(7226):1–2. doi: 10.1136/bmj.320.7226.1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  12. Jebb S. A., Cole T. J., Doman D., Murgatroyd P. R., Prentice A. M. Evaluation of the novel Tanita body-fat analyser to measure body composition by comparison with a four-compartment model. Br J Nutr. 2000 Feb;83(2):115–122. doi: 10.1017/s0007114500000155. [DOI] [PubMed] [Google Scholar]
  13. Leung S. S., Cole T. J., Tse L. Y., Lau J. T. Body mass index reference curves for Chinese children. Ann Hum Biol. 1998 Mar-Apr;25(2):169–174. doi: 10.1080/03014469800005542. [DOI] [PubMed] [Google Scholar]
  14. Nuñez C., Gallagher D., Visser M., Pi-Sunyer F. X., Wang Z., Heymsfield S. B. Bioimpedance analysis: evaluation of leg-to-leg system based on pressure contact footpad electrodes. Med Sci Sports Exerc. 1997 Apr;29(4):524–531. doi: 10.1097/00005768-199704000-00015. [DOI] [PubMed] [Google Scholar]
  15. Ogle G. D., Allen J. R., Humphries I. R., Lu P. W., Briody J. N., Morley K., Howman-Giles R., Cowell C. T. Body-composition assessment by dual-energy x-ray absorptiometry in subjects aged 4-26 y. Am J Clin Nutr. 1995 Apr;61(4):746–753. doi: 10.1093/ajcn/61.4.746. [DOI] [PubMed] [Google Scholar]
  16. Peng C. J., Chang C. M., Kuo S. E., Liu Y. J., Kuo H. C., Lin S. J. Analysis of anthropometric growth trends and prevalence of abnormal body status in Tainan elementary-school children. Acta Paediatr Taiwan. 1999 Nov-Dec;40(6):406–413. [PubMed] [Google Scholar]
  17. Schoeller D. A., Luke A. Bioelectrical impedance analysis prediction equations differ between African Americans and Caucasians, but it is not clear why. Ann N Y Acad Sci. 2000 May;904:225–226. doi: 10.1111/j.1749-6632.2000.tb06456.x. [DOI] [PubMed] [Google Scholar]
  18. Stewart S. P., Bramley P. N., Heighton R., Green J. H., Horsman A., Losowsky M. S., Smith M. A. Estimation of body composition from bioelectrical impedance of body segments: comparison with dual-energy X-ray absorptiometry. Br J Nutr. 1993 May;69(3):645–655. doi: 10.1079/bjn19930066. [DOI] [PubMed] [Google Scholar]
  19. Ward L. C., Heitmann B. L., Craig P., Stroud D., Azinge E. C., Jebb S., Cornish B. H., Swinburn B., O'Dea K., Rowley K. Association between ethnicity, body mass index, and bioelectrical impedance. Implications for the population specificity of prediction equations. Ann N Y Acad Sci. 2000 May;904:199–202. doi: 10.1111/j.1749-6632.2000.tb06449.x. [DOI] [PubMed] [Google Scholar]

Articles from Archives of Disease in Childhood are provided here courtesy of BMJ Publishing Group

RESOURCES