Skip to main content
. Author manuscript; available in PMC: 2007 Jan 3.
Published in final edited form as: Ultrasound Med Biol. 2006 Jun;32(6):857–874. doi: 10.1016/j.ultrasmedbio.2006.02.1428

Table 5.

Literature values for mechanical, ultrasonic and NMR properties of normal and abnormal breast tissue

Ultrasound
NMR
Tissue type Storage modulus Propagation speed (m/s) Atten. coeff. + freq. at 7 MHz (dB/cm/MHz) T1 (ms) T2 (ms)
Fibroadenoma - 1584 ± 27b - 1160 ± 336d 60 ± 11d,h
Cancer 93 ± 33a 1550 ± 35b 1.0 ± 0.2c 900 ± 225d 94 ± 48d
Glandular 33 ± 11a 1553 ± 35b 2.0 ± 0.7c 305e 49 ± 16d
Fat 19 ± 7a 1479 ± 32b 0.6 ± 0.1c 209 ± 17f 266 ± 45g 135 ± 16f 57 ± 3g

Values shown are means ± standard deviations, except for the case of T1 for the glandular parenchyma where no standard deviation was given. Unless otherwise stated, values result from measurements in vitro at 37°C or in vivo. Typical values of the storage modulus and attenuation coefficient for fibroadenomas were not found in the literature.

a

= Krouskop et al. (1998) at 1 Hz and at room temperature;

d

= Bottomley et al. (1987), Table 16, p 25. Values at 60 MHz, but temperature is not known;

g

= Nyman et al. (1986), 15 MHz;

h

= Small number of samples (5).