. 2002 Dec;88(6):622–626. doi: 10.1136/heart.88.6.622

Table 5.

Subgroup analysis of quantitative coronary angiography comparing optimal angioplasty with stent implantation versus direct stenting

Conventional angioplasty (+ stent) (n=47) Direct stenting (n=89) p Value
Balloon size (mm) 3.32 (0.31) 3.39 (0.27) 0.35
Balloon:artery ratio 1.00 (0.09) 1.05 (0.09) 0.068
Maximal balloon pressure (atm) 13.7 (1.2) 13.3 (1.6) 0.23
Number of stents 1.0 (0.0) 1.0 (0.0) 1
Length of stents (mm) 17.8 (3.6) 15.9 (1.7) 0.012
Reference diameter (mm)
    Before 3.30 (0.45) 3.22 (0.37) 0.46
    After 3.28 (0.44) 3.17 (0.33) 0.29
    At follow up 3.19 (0.44) 3.16 (0.40) 0.77
Minimal luminal diameter (mm)
    Before 0.47 (0.23) 0.62 (0.35) 0.035
    After 3.14 (0.61) 3.11 (0.54) 0.86
    At follow up 1.93 (0.98) 2.56 (0.86) 0.031
Stenosis (%)
    Before 85.3 (7.8) 80.5 (10.0) 0.022
    After 5.9 (8.4) 4.8 (9.4) 0.74
    At follow up 39.5 (30.5) 21.9 (22.8) 0.037
Gain (mm) 2.64 (0.64) 2.48 (0.64) 0.31
Loss (mm) 1.22 (0.90) 0.62 (0.69) 0.018
Net gain (mm) 1.53 (0.93) 1.99 (0.92) 0.15
Diameter stenosis >50% at follow up 13 (28.6%) 14 (15.7%) 0.011

Data are mean (SD).