. 2003 Aug;89(8):905–912. doi: 10.1136/heart.89.8.905

Table 3.

Coarctation: comparison of data before and after stent implantation and redilation

Pre-stent Post-stent Pre-redilatation Post-redilatation
Lumen diameter (mm) 5.7 (3.0– 8.0) 10.7 (7.0–13.3) 9.2 (6.3–11.1) 11.7 (9.8–14.1)
n=12; p=0.002 n=12; p=0.008 n=12; p=0.002
Gradient (mm Hg) 27 (13–75) 4.5 (0–20) 10.5 (0–20) 0.5 (0–15)
n=12; p=0.002 n=12; p=0.15 n=12; p=0.008
Diameter/DAo 0.49 (0.28–0.63) 0.81 (0.51–1.01) 0.66 (0.38–1.02) 0.85 (0.52–1.20)
n=12; p=0.002 n=12; p=0.01 n=12; p=0.005

Median values and ranges are given for each measurement. Data are compared with those in the preceding column. The number of measurements included in the comparison and the statistical significance of changes are indicated.

Diameter/DAo, ratio of coarctation lumen diameter indexed to the diameter of the descending aorta at the diaphragm.