Table 2.

Central corneal thickness measured by three methods.

Case Eye OCT (μm) Ultrasound (μm) Orbscan (μm)
1 OD 540 532 336
1 OS 571 550 481
2 OS 461 454 401
3 OD 552 540 573
3 OS 556 573 569
4 OD 496 486 408
4 OS 493 408 445
5 OS 512 d 377

OCT = optical coherence tomography

The mean difference between methods—OCT versus US: 6.4 ± 11.7 μm (P = .17, t test); Orbscan versus US: -67.5 ± 72.5 μm (P = .026, t test)