Skip to main content
. 2003 Aug;41(8):3661–3667. doi: 10.1128/JCM.41.8.3661-3667.2003

TABLE 5.

Sensitivity of the real-time RT-PCR multiplex assay

WNV standard (PFU) Results obtained with:
FAMa
VICb
CTc SD CT SD
8,000 17.20 0.15 17.54 0.16
800 20.75 0.08 21.07 0.08
80 24.37 0.12 24.62 0.14
8 27.66 0.12 28.34 0.14
0.8 30.55 0.24 31.97 0.24
0.08 40 0 40 0
0.008 40 0 40 0
a

WNV envelope-specific primer-probe set, with probe labeled with FAM fluorescent dye.

b

WNV NS1-specific primer-probe set, with probe labeled with VIC fluorescent dye.

c

Mean of CT values of 10 replicate samples.