. 2003 Aug;41(8):3661–3667. doi: 10.1128/JCM.41.8.3661-3667.2003

TABLE 6.

Reproducibility of results of the real-time RT-PCR multiplex assay

WNV standard (PFU) Results obtained with:
FAMa
VICb
CTc SD CVd CT SD CV
8,000 17.26 0.11 0.63 17.31 0.07 0.40
800 20.80 0.10 0.48 21.08 0.15 0.71
80 24.23 0.10 0.41 24.52 0.13 0.53
8 27.58 0.34 1.23 28.05 0.20 0.71
a

WNV envelope-specific primer-probe set, with probe labeled with FAM fluorescent dye.

b

WNV NS1-specific primer-probe set, with probe labeled with VIC fluorescent dye.

c

Mean of CT values of 10 replicate samples

d

CV, percentage of standard deviation relative to the mean.