. 1998 Jan 6;95(1):294–299. doi: 10.1073/pnas.95.1.294

Table 1.

Declining glucose characteristics

Location Onset time, min Decline rate, mg⋅dl−1⋅min−1 tminimum glucose, min Lag time, min
Blood samples 3.3  ± 0.5 6.8  ± 2.0 36.6  ± 7.2
Intravenous sensor 5.6  ± 1.7 7.0  ± 2.5 40.3  ± 5.9 3.7  ± 4.3
Subcutaneous sensor 8.9  ± 2.1 3.9  ± 1.3 61.2  ± 7.5 24.5  ± 6.8