Detection of dopamine and catechol using etched end-column electrochemical detection. Separations were performed in 770 nm i.d. capillaries with a 13-μm etch at the detection end. All values are the average of three separations on three different capillaries using three unique electrodes ± the standard deviation.

Dopamine Catechol
Peak Efficiency 60 000 ± 50 000 80000 ± 20000
Coulometric Efficiency 70 ± 10% 40 ± 20%
Detection limit [zmol] 400 ± 100 410 ± 80