TABLE 2.
Comparison of daptomycin E test MICs and NCCLS broth microdilution MICs according to the Mueller-Hinton agar test medium and E test formulation
| Type of M-Hb agar (calcium concn [μg/ml]) | E test format | n | No. of isolates with MIC differencec:
|
% EAa | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| −4 | −3 | −2 | −1 | Same | +1 | +2 | +3 | +4 | +5 | ||||
| BD (50) | Daptod only | 350 | 0 | 0 | 1 | 5 | 67 | 246 | 29 | 0 | 2 | 0 | 90.9 |
| BD (50) | Dapto + Ca | 350 | 1 | 10 | 118 | 198 | 22 | 1 | 0 | 0 | 63.1 | ||
| Remel (11-14) | Dapto only | 50 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 22 | 26 | 2 | 0 |
| Remel (11-14) | Dapto + Ca | 50 | 0 | 0 | 2 | 33 | 15 | 0 | 0 | 0 | 0 | 0 | 96 |
EA, essential agreement (percentage of MICs within 1 log2 dilution of the reference MIC).
M-H, Mueller-Hinton.
Difference values are log2 dilutions. Negative values indicate that the E test MIC is lower than the broth microdilution MIC.
Dapto, daptomycin.