. 2007 Jan 31;14(3):293–298. doi: 10.1128/CVI.00416-06

TABLE 2.

Raw data for instrument precision, with measurements by the Partec Cyflow SL_3 and the BD FACSCalibur/Sysmex XT1800i

Sample
Partec Cyflow SL_3
BD FACSCalibur/Sysmex XT1800i
CD4 lymphocyte count
% of CD4 lymphocytes in whole blood
CD4 lymphocyte count
% of CD4 lymphocytes in whole blood
Low Medium High Low Medium High Low Medium High Low Medium High
1 88 483 1,016 7 30 34 92 472 1,055 8 32 35
2 91 486 961 8 31 33 87 470 1,043 8 32 35
3 85 491 1,005 7 30 34 89 482 1,070 8 33 35
4 94 463 1,018 8 29 34 92 506 1,073 8 34 36
5 94 470 1,067 8 30 35 87 499 1,064 8 34 35
6 93 493 1,037 8 30 34 92 466 1,059 8 31 35
7 92 484 1,034 8 30 34 105 456 1,084 9 31 36
8 86 476 996 7 29 34 84 479 1,036 7 32 34
9 96 466 986 8 29 33 96 464 1,057 8 31 35
10 92 469 1,012 8 29 34 94 478 1,057 8 32 35
Mean 91 478 1,013 7.7 29.7 33.9 92 477 1,060 8.0 32.2 35.1
SD 4 11 29 0.5 0.7 0.6 6 15 14 0.5 1.1 0.6
CV 3.99 2.25 2.90 6.27 2.27 1.67 6.41 3.25 1.32 5.89 3.53 1.62
a

Single large preparations were made for each of the three samples and analyzed 10 consecutive times. SD, standard deviation.