TABLE 3.

Performance parameters of BacT/ALERT 3D system and PZase assay after resolution of the discordant results

Performance parameter % witha:
BacT/ALERT 3D system PZase assay
Sensitivity 100 (35/35) 82.85 (29/35)
Specificity 98.61 (71/72) 95.83 (69/72)
Accuracy 99.06 (106/107) 91.58 (98/107)
Predictive value for resistance (PPV)b 97.22 (35/36) 90.63 (29/32)
Predictive value for susceptibility (NPV)c 100 (71/71) 92.0 (69/75)
a

Values in parentheses are the number of isolates; the first number represents the number of isolates detected for each parameter, and the second number represents the total number of isolates tested.

b

PPV, positive predictive value.

c

NPV, negative predictive value.