Operational parameters and resulting output powers and efficiencies of MIT gyrotrons for DNP/NMR experiments.

140 GHz 250 GHz 460 GHz

Operation voltage, Vo (kV) 12 12 12
Beam current, Io (mA) 25 25 135
Operating mode TE031 TE032 TE061
Tube output mode TE01 HE11 HE11
Magnetic field, Bo (T) 5.1 9.0 8.4
Cyclotron harmonic number 1 1 2
Output power (W) 14 7 8