Table 2.

Effect of Filters on Quality of MR Spectra*

RMS noisea Linewidthb Poor fitc
No filter 7.2 ± 1.7 5.02 ± 2.19 5.57%
Apodization 4.5 ± 1.5 5.90 ± 2.20 5.82%
DSM 3.4 ± 1.5 5.27 ± 2.17 3.94%
*

Results from 44,695 MR spectra from 22 control subjects using no spectral filters, a Gaussian apodization filter of 1.3 Hz width, or DSM.

a

RMS: root mean square.

b

Linewidth in Hz.

c

Percentage of spectra in MRSI that did not meet following criteria: 1) linewidth between 1.9 and 11.1 Hz; 2) fitted resonance difference between MI and Cr between 0.43 and 0.63 ppm; and 3) fitted resonance difference between NAA and Cr between 0.97 and 1.05 ppm.