Skip to main content
. 2006 Jun;90(6):709–712. doi: 10.1136/bjo.2005.086942

Table 2 Treatment powers and MPE multiplication factors*.

Patient number
1 2 3 4 5 6
Power needed for a visible CW lesion 300 mW 300 mW 250 mW 300 mW 300 mW 300 mW
Energy needed for a visible CW lesion 60 mJ 60 mJ 50 mJ 60 mJ 60 mJ 60 mJ
CW MPE multiplication factor† 36× 36× 30× 36× 36× 36×
Power needed for a visible MP lesion 1800 mW 1800 mW 1800 mW 1800 mW 1800 mW
Energy needed for a visible MP lesion 54 mJ‡ 54 mJ 54 mJ 54 mJ 54 mJ
MP MPE multiplication factor† 133× 133× 133× 133× 133×

*Abbreviations CW and MP in this table connote continuous wave and micropulse exposures, respectively.

†The number of times that delivered corneal energy exceeds ANSI Z136.1‐2000 maximum permissible exposure (MPE)13 for corneal energy. For a 125 μm retinal laser spot size diameter, 810 nm, 0.2 second envelope, 500 Hz exposure of 100, 0.3 ms micropulses, CA = 1.660 and CE = 4.902. CA is the correction factor used to increase MPE values in the near infrared (IR‐A) due to decreased RPE absorption in IR‐A.13 CE is the correction factor used to calculate the ocular extended source MPE from the small source MPE when the laser source subtends a visual angle greater than αmin (the apparent visual angle which divides small source from extended source viewing).13

‡(power)×(duration)×(duty factor) = (1800 mW)×(0.2 s)×(0.15) = 54 mJ

¶As noted in table 1, micropulse lesions were not demonstrable in this patient even with power settings 6× MPE.