Table 1 Comparison of Sysmex UF‐100 flow cytometer and bacterial culture results using variable bacterial counts as the cut‐off values.

Cut‐off value (number of bacteria/μl)
8000 7000 6000 5000 4 000 3 000 2 000 1 000
Number of culture positive/Sysmex positive 265 268 269 275 281 290 297 304
Number of culture negative/Sysmex positive 162 176 198 223 263 316 420 591
Number of culture negative/Sysmex negative 537 523 501 476 436 383 279 108
Number of culture positive/Sysmex negative 41 38 37 31 25 16 9 2
Sensitivity 86 87 88 90 92 92 97 99
Specificity 77 75 72 68 62 55 40 15
Positive predictive value 62 60 56 55 52 47 41 34
Negative Predictive value 93 93 93 94 94 96 97 98
Number of urine samples flagged for culture, total (%) 427 (42) 444 (44) 467 (46) 498 (49) 544 (54) 606 (60) 717 (71) 895 (89)