Skip to main content
. 2007 Feb 7;14(4):412–419. doi: 10.1128/CVI.00372-06

TABLE 2.

Biases among cell counts obtained by our cell enumeration system (SP ICM), TruCount, and DP data for different cell populations

Comparisona Bias from Bland-Altman plot (%) for the following cell type (antibody)b:
Leukocytes (CD45) T lymphocytes (CD3) B lymphocytes (CD19) T lymphocytes (CD4)
(TruC − ICM)/avg (%) vs avg 10.1 10.0 0.3 −37.5
(DP − ICM)/avg (%) vs avg −12.0 −12.9 −20.1 −56.4
(TruC − DP)/avg (%) vs avg 22.0 22.8 20.5 20.0
a

Expressed as the difference between the results by two methods divided by the average of those results, expressed as a percentage, versus the average of the results obtained by the two methods. TruC, TruCount.

b

The number of patients was 50 for leukocytes (CD45), 52 for CD3+ T lymphocytes, 42 for B lymphocytes (CD19), and 45 for CD4+ T lymphocytes.