Skip to main content
. 2007 Apr 25;37(6):515–529. doi: 10.1007/s00247-007-0434-0

Table 3.

Volumetric CT chest scanning parameters according to child’s weight when routine and Combiscan protocols are performed

  <15 kg 15–24 kg 25–34 kg 35–44 kg 45–55 kg
Volume Combi Volume Combi Volume Combi Volume Combi Volume Combi
kVp 100 100 100 100 100 100 100 100 100 100
Effective mAs 20 20 25 25 35 35 55 55 75 75
Collimation (mm) 1.5 0.75 1.5 0.75 1.5 0.75 1.5 0.75 1.5 0.75
Scan slice width (mm) 5 5 5 5 5 5 5 5 8 8
Table feed (mm) 24 12 24 12 24 12 24 12 24 12
Scan time (s) 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5
Calculated Effective Dose (mSv)–CT EXPO 0.9 1.0 1.13 1.31 1.58 1.75 2.48 2.75 3.38 3.75