Skip to main content
. 2007 May 23;7:45. doi: 10.1186/1471-2334-7-45

Table 2.

Workflow of reprocessing in the washer/disinfector and further processing

Step Parameters
Flushing 25°C, 2 min, RO (reverse osmosis) water
Basic cleaning and disinfection (NaOH) pH 11, RO water, 50°C/5 min
Rinsing RO water 25°C/1 min
Acidic cleaning to remove mineral stains of calcium carbonate phosphoric acid, pH 2, RO water, 50°C/5 min
Final rinsing RO water 25°C/1 min
Thermic disinfection 95°C/10 min, RO water
Check for membrane integrity resistance time over 1 min
Drying with sterile air 115°C
Packing and documentation sterile boxes