False-positive results in multiple immunometric assays due to an interference that changed over time8

Analyte /Assay system Untreated sample Patient sample pre-treated with patient's E. coli isolatea
TNI Dade Dimension RxL (μg/L) 235 <2
TNI Dade Stratus II (μg/L) <0.4 Not done
AFP Abbott AxSYM (μg/L) 41 5
CA-125 Abbott AxSYM (kU/L) 225 14
hCG Abbott AxSYM (IU/L) 17 <5
TSH Abbott AxSYM (mU/L) 14 1.4
a

Incubation of sample with patient's E. coli isolate for 16–18 h prior to assay