Skip to main content
. 2007 May 16;45(7):2105–2109. doi: 10.1128/JCM.00210-07

TABLE 2.

Comparison of technical parameters among Nanogen GNEN, Genaco ResPlex II, and monoplex TaqMan methods for detection of RNA virus respiratory pathogens

Parameter GNEN ResPlex II TaqMan
Amplification platform Multiplex RT-PCR Multiplex RT-PCR Monoplex RT-PCR
Detection format Nanochip (solid chip) Luminex (liquid chip) TaqMan (real time)
Pathogens covered Flu-A, Flu-B, PIV-1, PIV-2, PIV-3, RSV Flu-A, Flu-B, PIV-1, PIV-2, PIV-3, PIV-4, RSV-A, RSV-B, hMPV, RhV, EnV, SARS CoV Flu-A, Flu-B, PIV-1, PIV-2, PIV-3, RSV, hMPV
No. of reactions needed 1 1 7
Hands-on time (min)a 70 55 135
Detection throughput (no. of tests/run) 80 96 13.5
Test turnaround time (h)b 9 6 4
a

Hands-on and turnaround times include time needed for specimen processing and data analysis.

b

Times are based on a full run, including specimen processing, target amplification, and product identification.