Skip to main content
. 2007 Apr 4;45(6):1723–1727. doi: 10.1128/JCM.02558-06

TABLE 2.

Reproducibility of the Abbott test using plasma samples

Nominal concn (log10 copies/ml) No. of samples tested Mean viral load (log10 copies/ml) SD (log10 copies/ml) % CVa
6.0 6 5.88 0.19 3.2
5.0 16 4.79 0.24 5.0
4.0 16 3.91 0.21 5.3
3.0 16 2.76 0.24 8.7
2.3b 19 1.93 0.45 23.3
2.0b 18 1.90 0.46 24.2
a

CV, coefficient of variation.

b

Only positive samples were included.