. 2007 Apr 4;45(6):1712–1717. doi: 10.1128/JCM.02385-06

TABLE 2.

Comparison of the Abbott RealTime HCV and HIV-1 viral load assays with comparator assays

Abbott RealTime assay Assay used versus Abbott RealTime assay No. of samples r2a Mean differenceb Regression (y = ax + b)a
% of samples with difference
a b Comparator < Abbottc Abbott < comparatord
HCV LcX 92 0.89 −0.35 0.95 0.54 2.2 3.2
COBAS Amplicor 146 0.81 −0.08 0.75 1.41 2.7 2.0
COBAS TaqMan 161 0.83 −0.25 0.84 1.02 4.3 4.3
bDNA 36 0.98 0.06 0.93 0.12 0 0
HIV LcX 128 0.82 −0.11 1.00 0.14 3.1 1.6
COBAS Amplicor 115 0.84 −0.14 0.87 0.61 0.9 0.9
COBAS TaqMan 55 0.76 −0.24 0.93 0.62 1.8 1.8
bDNA 168 0.90 −0.14 0.93 0.38 0 1.8
a

Spearman regression coefficient and equation of regression analysis.

b

The mean difference was calculated as follows: mean log10 Abbott RealTime assay − mean log10 comparator results.

c

Percentage of samples with a mean difference 1 log10 higher in the Abbott RealTime assay than in the comparator assay.

d

Percentage of samples with a mean difference 1 log10 higher in the comparator assay than in the Abbott RealTime assay.