Table 1.

Reaction volume and heating/cooling rates of PCR chips

Reaction volume (nl) Heating rate (°C/s) Cooling rate (°C/s) Amplification time (s)/Cycle number Amplicon length (bp) References
Chamber PCR
Small volume
0.45 240 (49)
33 (78)
40 ∼6000 or ∼3720/40 74 (17)
100 175 125 340/40 82 (80)
200 10–50 5 600/30 150/1382 (77)
240 >20 >20 1320/35 690 (32)
280 4.8/7.8 5.0 708/30 278 (26)
380 >15 >10 <1620/30 (31,35)
420 3 2 (39)
550 660/30 211 (27)
900–7000 6.5 1.95 (54)
1000 44 20 208 (79)
1170 522/30 300 (13)
1750 4.5 4.5 322 (53)
2000 11 3.2 1350/30 330 (11)
Large volume
∼3500 11 2.7 ∼1680/30 209 (14)
3600 ∼10 ∼4.6 ∼2400/30 183 (51)
8000 5 2.67 1740/25 305 (62,63)
10 000 305 (57)
10 000 ∼20 ∼10 900/20 240 (40)
10 000 ∼20 ∼10 840/20 273 (47)
10 000 ∼20 ∼10 1584/30 273 (45)
15 000 38 7.9 3300/25 511 (89)
∼20 000 11.5 5.5 <900/20 273 (38)
25 000 8 8 1680/30 244 (70)
37 500 3.1 3.1 2100/40 544 (37)
50 000 ∼20 ∼10 511 (48)
Continuous-flow PCR
(480 ∼ 1800)/30 430 (44)
500 1086/27 500 (56)
1000 ∼900/35 (23)
∼6000 2100/25 145 (41)
4500 ∼12 ∼12 2280/23 240 (64)
8000 300/>40 372 (65)