Skip to main content
. 2003 Sep;41(9):4016–4021. doi: 10.1128/JCM.41.9.4016-4021.2003

TABLE 2.

Comparison of Ct values in monoassay and multiplex assay after real-time amplification of a dilution series of an L. pneumophila serogroup 1 standard

CFU/ml Ct
Monoassay
Multiplex assay
L. pneumophila Legionella spp. L. pneumophila Legionella spp.
100,000 27.2 27.7 27.4 27.8
10,000 31.2 30.6 31.6 32.8
1,000 34.3 34.5 35.2 35.5
100 38.5 38.3 38.6 39.4
10 43.4 42.4 43.2 44.7