Skip to main content
. Author manuscript; available in PMC: 2008 Mar 1.
Published in final edited form as: Ophthalmology. 2006 Nov 30;114(3):466–471. doi: 10.1016/j.ophtha.2006.08.022

Table 2.

Area under the Receiver Operating Characteristics (95% Confidence Interval) for Discrimination between Healthy and Glaucomatous Eyes Using Heidelberg Retina Tomograph (HRT) 3 Stereometric Parameters

Characteristic Numerical Output Ordinal Output
HRT 3
 Linear cup/disc ratio 0.897 (0.836–0.958) 0.693 (0.623–0.763)
 Cup shape measure 0.871 (0.809–0.932) 0.664(0.596–0.731)
 Rim area 0.839 (0.759–0.920) 0.815 (0.744–0.886)
 Rim volume 0.825 (0.745–0.906) 0.753 (0.677–0.828)
 Mean RNFL thickness 0.790 (0.702–0.877) 0.623 (0.560–0.686)
 Height variation contour 0.615 (0.504–0.726) 0.530 (0.497–0.563)
GPS
 Horizontal RNFL curvature 0.905 (0.844–0.966) NA
 GPS global 0.899(0.837–0.961) 0.862 (0.794–0.929)
 Cup size 0.768 (0.678–0.858 NA
 Vertical RNFL curvature 0.731 (0.636–0.731) NA
 Rim steepness 0.686 (0.586–0.786) NA
 Cup depth 0.560 (0.453–0.666) NA

GPS = glaucoma probability score; NA = not applicable; RNFL = retinal nerve fiber layer.