Table 1.

Performance requirements for ideal HTS assays

Parameter Performance standard
S : B > 3
separation band or assay window 1 SD between the mean maximum value - 3 SD and the minimum value + 3 SD
Z-factors greater than 0.5
intra-plate variability < 20% COV
inter-plate variability each plate within 20% of the mean of the 2 plates
day-to-day variability < 20% COV between plate pairs

SD, standard deviation

COV, coefficient of variance